FLUKE 80T-IR PDF

80T-IR – read user manual online or download in PDF format. Pages in total: Find great deals for Fluke 80t-ir Temperature Probe 80TIR. Shop with confidence on eBay!. Find great deals for Fluke Infrared Temperature Probe M# 80t-ir. Shop with confidence on eBay!.

Author: Shakabei Gut
Country: Suriname
Language: English (Spanish)
Genre: Health and Food
Published (Last): 15 October 2015
Pages: 427
PDF File Size: 17.84 Mb
ePub File Size: 6.81 Mb
ISBN: 582-5-44619-531-9
Downloads: 12021
Price: Free* [*Free Regsitration Required]
Uploader: Akinogar

Adjust the setpoint of the Blackbody to output each of the temperatures in Table 2 at 0. Add to a parts list. With the UUT powered on, check that the battery voltage is at least 7. The product does not contain any of the restricted substances in concentrations and applications banned by the Directive, and for components, the product is capable of being worked on at the higher temperatures required by lead—free soldering.

Fluke 80T-IR User Manual

The Manufacturers disclaim all warranties including implied warranties of merchantability or fitness for a particular purpose and are not liable for any damages arising from your use of or your inability to use the Information downloaded from this website. Recommended Equipment Table 1 lists the equipment required to perform the calibration verification procedure. Search the history of over billion web pages on the Internet.

  AMORETTI SPENSER PDF

May substitute less accurate blackbody if using in conjunction with transfer standard of equivalent accuracy, i.

Fluke 80t-ir Infrared Temperature Probe 80tk Thermocouple Module

Please select an existing parts list. Power up the UUT and allow it to stabilize for five minutes. Please enter a message. Place the UUT aperature three inches from the center and perpendicular to both axes 80t–ir the Blackbody plate.

Fluke 80T-IR Non-Contact Temperature Probe

Save to an existing parts list Save to a new parts list. Sending feedback, please wait Refer to Figure 1 for equipment configuration. Verification Points Model Blackbody Temperature at 0. Save to parts list Save to parts list. You have chosen to save the following item to a parts list:. If your instrument fails to pass any performance specifications, return it to the Eluke Eactory or Service Center for repair.

The foregoing information relates to product sold on, or after, the date shown below. Record the reading displayed on the DMM. You agree that the Information as provided here through an intermediary may not be error-free, accurate or up-to-date and that it is not advice.

  ISO 13818-2 PDF

In between measurements move the UUT to an environment not affected by the heat source. Thank you for your feedback. The Manufacturers reserve fluk right to change this Information at any time without notice. Repeat steps 1 through 5 for the remaining verification points called out in Table 2. Accuracy figures are valid for a period of one year.

The product detailed below complies with the specifications published by RS Components. Replace the battery if necessary.

Reactivate by switching the UUT off and on again. Save this item to a new parts list. Set the DMM to the 3.